8.0 Maintenance and Service

All Boekel laboratory products are designed to comply with

IEC1010-1. No routine maintenance is required.

8.1 Cleaning

Disengage power cord prior to cleaning. If a spill occurs, use

appropriate clean-up procedures as required for radiation or

biohazard control. The outer casing can be cleaned with a cloth

dampened with water. Do not immerse the unit in water.

8.2 Replacement of Fuses

There are two supply fuses located in the fuse drawer. To change

the fuses:

Turn power switch to the off (O) position

Disconnect the unit from the power supply

Remove the line cord from the power entry module on the back

of the unit

Pull back on the fuse drawer catch

Pull out the fuse drawer

Check and replace with the correct fuses if necessary. The

fuses should be 5mm x 20mm quick acting, rated 250V.

Model 240000 (115 V AC):

2AF

Model 240000-2 (230 V AC):

1AF

Push the fuse drawer back in. Reconnect unit to power supply.

9.0 Warranty

When used in laboratory conditions and according to these

operating instructions, Boekel warrants this product to be free of

defective material and workmanship for a period of two years from

the date of manufacture. The liability of Boekel for any defective

equipment during the warranty period shall be limited to the repair

of such equipment or replacement thereof without charge for parts

or labor.

10.0 Service

It is required to obtain a Returned Material Authorization (RMA)

number before any Boekel products are returned for any reason. A

Decontamination Certificate must be completed, signed by the

user, and returned to Boekel Scientific prior to receiving the RMA

number. Please be sure to mark the outside of the returned goods

package with this RMA number to ensure prompt handling