2 preparing the unit, 3 loading the unit, 4 removing the samples – Boekel Scientific 240000-2 Slide Moat User Manual

Page 6: 5 setting the temperature, 0 technical specifications

background image

5

The Power Switch controls power to the unit.
The Heat Indicator is on continuously while the oven is heating up.

As the required temperature is approached, it starts to flash. When

the unit is controlling at the set temperature, the heater lamp

flashes intermittently.
The Temperature Display shows the chamber temperature in

degrees Celsius.
4.2 Preparing the unit

Switch the unit on (I) and set the PID Controller to the desired

operating temperature. Although the unit heats rapidly, you may

want to first pre-heat the unit to the desired operating temperature

before incubating samples. If a humid chamber is desired, fill each

of the eight humidity wells with 0.1ml of distilled water.

4.3 Loading the unit

Place slide samples on the land areas of the heat plate. Place the

tempered glass lid on the gasketed perimeter of the heat chamber.

Secure the glass lid by gently depressing the lid one corner at a

time and rotating the plastic keeper over the lid edge. As you

release the glass lid it will secure itself against the keeper. If water is

being used, you will notice a slight fogging above each well as the

unit generates and maintains the humid environment.

4.4 Removing the samples

Caution! The lid may be hot. Make sure the lid is cool before

attempting to remove keepers. Release lid keepers by again

depressing the lid at each corner and rotating the keeper off the

lid. Remove the glass lid. Caution! The heating plate may be hot.

Allow the heating plate and slides to cool before removing the

slides.

4.5 Setting the Temperature

The temperature controller has three buttons. When the button on

the left, ‘*’, is depressed, the set temperature is displayed. When

the left button ‘*’ is depressed simultaneously with the middle

button ‘’, the set temperature value is lowered. When the left

button ‘*’ is depressed simultaneously with the right button ‘’, the

set temperature value is raised. When all buttons are released, the

actual chamber temperature is displayed.

5.0 Technical Specifications

This equipment is for indoor use and will meet its performance

figures within the ambient temperature range of 10°C to 35°C, with

maximum relative humidity of 80% non-condensing. Installation

category II (transient voltages). Pollution degree 2 in accordance

with IEC 664. For operation at altitudes of up to 6500 feet.