0 fault diagnosis, 0 accessories – Boekel Scientific 240000-2 Slide Moat User Manual
Page 7

6
Temperature Range
Ambient + 5°C to 100°C
Temperature Setting Range 10.0°C to 100.0°C
Stability
±0.2°C throughout operating temperatures
Uniformity
±0.35°C at 37°C; ±0.7°C at 65°C
Temperature Display
Resolution
0.1°C
Supply Ratings
Model 240000: 115 VAC, 1.74 A, 60 Hz
Model 240000-2: 230 VAC, 0.87 A, 50/60 Hz
Power Rating
Model 240000: 200 W
Model 240000-2: 200 W
Heating Rate
Ambient to 65°C in 20 minutes
6.0 Fault Diagnosis
Symptom
Possible Cause
Action Required
Unit does not operate
a. Unit not switched on
b. Unit not plugged
into power supply
c. Fuses blown
d. Power supply failure
a. Switch on
b. Plug in, switch on
c. Replace fuses (see
section 8.2)
d. Check that other
electrical appliances
on the same circuit
are working
Chamber temperature
does not rise when
expected
a. Actual temperature
is higher than set
temperature
b. Temperature
control circuit fault
a. Check set
temperature
b. Have unit checked by
a competent service
person
Temperature continues
to rise when not
expected
a. Actual temperature
is lower than set
temperature
b. Temperature
control circuit fault
a. Check set
temperature
b. Have unit checked by
a competent service
person
7.0 Accessories
Part Number
Description
C1303205
Light Shield Cover