8. Maintenance and Service

All Boekel laboratory products are designed to comply with IEC1010-1. No routine
maintenance is required.

8.1

Cleaning

Disengage power cord prior to cleaning. The outer casing may be cleaned with water and
a damp cloth. Do not submerge or immerse the Round Tissue Bath in water. Before
using any cleaning or decontamination method except those recommended by the
manufacturer, users should check that the proposed method would not damage the
equipment.

8.2

Replacement of Fuses

There are two supply fuses located inside the Round Tissue Bath. Fuse replacement must
be done only by a qualified service person.

• Disconnect the unit from the power supply.

• Remove the screws from the bottom of the unit and lift the pan out of the

body.

• Check and replace with the correct fuses if necessary. The fuses must be ¼”

X 1 ¼” quick acting, rated 250V. Replacement fuses are available from
Boekel.

Model 14792:

-2AF

(Boekel part number 902-0068)

Model 14792-2-2:

-1AF

(Boekel part number 902-0101)

• Carefully replace the pan into the body. Make sure that no wires are pinched

when inserting the pan into the body.

• Replace the screws in the bottom of the unit.

• Reconnect unit to the power supply.

9. Warranty

When used in laboratory conditions and according to these operating instructions, Boekel
warrants this product to be free of defective material and workmanship for a period of two
years from the date of manufacture. The liability of Boekel for any defective equipment
during the warranty period shall be limited to the repair of such equipment or replacement
thereof without charge for parts or labor.

10. Service

A Returned Goods Authorization (RGA) number must be obtained before any Boekel
products are returned for any reason. A Decontamination Notice must be completed,
signed by the user, and returned to Boekel Scientific prior to receiving the RGA number.
Please be sure to mark the outside of the return goods package with this RGA number to
ensure prompt handling.