2 filling the bath, Fault diagnosis, Technical specifications – Boekel Scientific 14792-2 Round Tissue Bath User Manual

Page 5

background image

4

4.2

Filling the Bath

Fill the bath with water. Always keep at least 3/4” (1.9cm) of water in the bath. Do not
let the bath run dry as heater damage may result.

5. Fault Diagnosis


Symptom

Possible Cause

Action Required

1. Unit does not operate

a. Unit not switched on
b. Unit not plugged into

power supply

c. Fuses blown
d. Power supply failure

a. Switch on
b. Plug in, switch on

c. Replace fuses per 8.2
d. Check that other electrical

appliances on the same
circuit are working

2. Chamber temperature does

not rise when expected

a. Actual temperature is

higher than set
temperature

b. Temperature control

circuit fault

a. Check set temperature


b. Have unit checked by

competent person

3. Temperature continues to

rise when not expected

a. Actual temperature is

lower than set temperature

b. Temperature control

circuit fault

a. Check set temperature

b. Have unit checked by

competent person

7. Technical Specifications


This equipment is intended for indoor use and will meet its performance figures within the
ambient temperature range of 10

°C to 35°C, with maximum relative humidity of 80% (non-

condensing). Installation Category II (transient voltages). Pollution Degree 2 in accordance
with IEC 664. Suitable for operation at altitudes of up to 6500 feet.

Specifications:

Temperature Range:

(Ambient +10

°C) to 65°C

Setting Range:

Ambient to approximately 70

°C

Supply Voltage Range:

115V +/- 10%, 50/60 Hz

230V +/- 10%, 50/60 Hz

Power Rating:

Model 14792: 190W

Model 14792-2: 190W

Heating Rate:

Ambient to 65

°C within 60 minutes