2. function list – A&D Weighing SE-150KAL User Manual
Page 23
21
10-2. Function list
Item
Parameter
Description
0
Auto power-off disabled
Auto power-off
function
poff
1
Auto power-off enabled
Turns the power OFF
automatically
0
Normal (1/3,000)
Weight display
resolution
r5eo
1
High (1/6,000 or 1/7,500)
Changes “d”,
minimum display
0
2400 bps
1
4800 bps
Serial interface
Baud rate
bp5
2
9600 bps
0
Stream mode
Command mode
1
Command mode only
2
Print switch mode
Command mode
3
Auto-print mode +/- data
Command mode
Serial interface
Data output
mode
prt
4
Auto-print mode + data
Command mode
0
No reply except “Q” command
Serial interface
Response
aCk
1
Reply to commands
0
Comparator disabled
1
Compares all data
2
Compares all stable data
3
Compares data of > +4d or < -4d
4
Compares stable data of > +4d or < -4d
5
Compares data of > +5d
Comparator
mode
Cp
6
Compares stable data of > +5d
Comparison conditions
d = minimum display
0
Weak / fast response
1
Normal / normal response
sStability
Response speed
Cond
2
Strong / slow response
0
Zero tracking function disabled
Zero tracking
trc
1
Zero tracking function enabled
Tracks the zero drift
0
Dot (.)
Decimal point
dp
1
Comma (,)
Selects a decimal
point type
.
no Initialization disabled
Initialization
Clrf
go Initialization enabled
Restores the function
settings to the factory
setting values.
Factory setting
d = minimum display, the minimum mass that can be weighed
Even the counting mode uses “d” for judgment.