10-2. GLP report
To print the GLP report to the AD-8121B, select the function setting “info 1” and
“pUse 1” for the balance, and use MODE 3 of the printer.
To transmit the GLP report to a personal computer, select the function setting “info
2” and “pUse 0”.
The serial number attached to the lateral side of the balance may have a form
“#Q1234567” (# = I, E and so on). In this case, the first letter “#” is ignored as a
GLP data and output data has a form “Q1234567”.
Calibration report
This function is not available after sealing the calibration (CAL) switch.
1. Perform calibration according to “7-1. Calibration
using a weight”.
2. end appears when the calibration has
been completed.
3. glp is displayed and calibration report is
output.
4. end appears again. Remove the weight
and press the MODE key to return to the
weighing mode.
AD-8121 format “info 1”
General format “info 2”
‘
: Space,
ASCII
20h
CR: Carriage return, ASCII 0Dh
LF: Line feed, ASCII 0Dh
end
glp
To output
calibration report
Operation of Calibration
(See “7.CALIBRATION”)
end
Returns to the weighing mode
`~~~~~~~~~~
A
~
&
~
D
MODEL
~~~
EK-1200i
S/N
~~~~~
Q1234567
ID
~~~~~~~~
ABCDEF
DATE
TIME
CALIBRATED(EXT.)
CAL.WEIGHT
~~~~~~
+1000.0
~~
g
SIGNATURE
---------------
~
A & D
MODEL EK-1200i
S/N Q1234567
ID ABCDEF
DATE 2011/11/14
TIME 02:53:21
CALIBRATED(EXT.)
CAL.WEIGHT
+1000.0 g
SIGNATURE
- - - - - - - -
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration
executed
Calibration weight
Column for
signature