A&D Weighing BM-500 Addendum User Manual
Page 2
2
External Control of the Ion Function (Static Charge Eliminator)
□
To assign the Ion function to the MODE key of the AD-8922A remote controller, select the
“
ionfnc
” (Static elimination) settings and change the “
8922
” (AD-8922A control) setting.
□
To assign the Ion function when using both the AX-T-314A-S external key plug and AX-SW128
foot switch (both sold separately), select the “
ionfnc
” (Static elimination) settings and change
the “
e-5
” (external input control) setting.
Ion function control settings (added after the setting “ion”)
■
0
3 seconds
1
10 seconds
ion
Neutralization time
2
Manual operation (max. 10 min.)
■
off
Do not control the Ion function with the
AD-8922A
8922
AD-8922A control
on
Assign the Ion function to the MODE key of
the AD-8922A.
■
0
1. PRINT key, 2. RE-ZERO key
1
1. PRINT key, 2. ION key
ionfnc
Static
eliminator
e-5S
External input control
(See the figure below
for terminals)
2
1. ION key, 2. RE-ZERO key
■
Factory setting
Terminal 1
Terminal 2
GND
*External key plug
AX-T-314A-S
(Sold separately)
New