Kistler-Morse KM SVS2000 User Manual
Page 7

j. When the desired baud rate is displayed, press the ‘ENTER’ key. The display reads: BAUD RATE
k. Press the ‘GROSS’ key to exit the PLC Menu Tree.
If you started this procedure with the SVS2000 Field Firmware Upgrade then you will need to go back to those
instructions now in order to restore the calibration/setup parameters.
FOR ETHERNET/IP
TM
The Ethernet/IP
TM
Module is shipped from the factory configured for an IP address of 192.168.1.2 with a Subnet
Mask of 255.255.255.0 and the Default Gateway of 0.0.0.0.
d. Press the ‘DOWN’ arrow key until the Ethernet/IP
TM
menu appears. Press the ‘ENTER’ key. The
display reads: IP Address.
e. Press the ‘ENTER’ key. The default IP Address of 192.168.1.2 is shown and allows for editing.
Type the new IP address using the keypad. The ‘ENTER’ key will advance through each of the
four (4) octets, and the last octet entered will store the value; the display will show IP Address.
f. Press the ‘UP’ arrow key to access the Subnet Mask. Press the ‘ENTER’ key. The Subnet
Mask of 255.255.255.0 is shown and allows editing. Type the new Subnet Mask using
the keypad. The ‘ENTER’ key will advance through each of the four (4) octets, and the last octet
entered will store the value; the display will show the Subnet Mask.
g. Press the ‘UP’ arrow key to access the Default Gateway. Press the ‘ENTER’ key. The Default
Gateway of 0.0.0.0 is shown and allows editing. Type the new Default Gateway using
the keypad. The ‘ENTER’ key will advance through each of the four (4) octets, and the last octet
entered will store the value; the display will show the Default Gateway.
4. Configure the PLC to match the selections made to the SVS2000.
If you started this procedure with the SVS2000 Field Firmware Upgrade then you will need to go back to those
instructions now in order to restore the calibration/setup parameters.
3
www.kistlermorse.com
97-1501-12 Rev. B