8 Troubleshooting - Error messages
991-0251 Rev. 1.0
Feb-2013
19
8 Troubleshooting - Error messages
Under certain conditions the NucleoCounter® NC-250™ will display error messages during
operation. Corrective actions are suggested below, but if they do not correct the errors,
contact ChemoMetec A/S or the local distributor.
Under certain conditions the NucleoView™ NC-250™ software will display error messages
during operation. Corrective actions can be found in the NucleoView™ NC-250™ software
Users Guide.
8.1
Required hard disk space for data storage.
The CM files stored on the hard disk can vary from 1 5MB up to in some cases more than
50MB. Therefore it is important from time to time to check whethe r enough free space is
present on the hard disk. ChemoMetec recommends a minimum of 10GBs free space on
the hard disk.
It is recommended to make a periodic archiving of CM files to external storage. Those
archived CM files can then be removed from the har d disk of the PC.
The periodic archiving can be made on either the creation date of the CM file, or when the
free disk space is below a certain amount, or when the disk space used for CM files
exceeds a certain amount. Contact the local IT Administrator in order to get information
about procedures for external data storing.
A good example of the archiving procedure could be based upon:
CM files older than 3 months are archived
CM files uses more than 50GBs of disk space, archive the oldest CM files until le ss
than 20GB is used
When less than 10 GBs free space, archive the oldest CM files until less than 20GB
is used.
Refer to the NucleoView™ NC-250™ Software User’s Guide
5
for additional information.
8.2
Power-on failure
If the NucleoCounter® NC-250™ cannot be turned on, please verify that the power supply
is connected to the NucleoCounter® NC-250™ and a working mains plug.
If the NucleoCounter® NC-250™ still does not turn on, inspect and/or replace the fuse as
described below.
The fuse holder is located at the rear side above the power plug, see Figure 13
5
Locate the file ”Install_Guide.html” on the USB Flash drive and follow the directions.