CT-2000 Constant Temperature Bath

Revision 1.0g—February, 2012; CANNON

®

Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

If the CT-2000 was previously configured to operate in a polled (RS-485)
environment, the following will appear on the LCD:

COMMUNICATION MODE

1. Polled

2. Point to Point

(Reports Disabled):1

The last line on the LCD indicates that timed reports are disabled and the
present mode is polled.

In this example, to operate in a point-to-point (RS-232) environment with
or without timed interval reports, you would then press:

[2] [ENTER]

NOTE

To configure reports, see page 20.

If the CT-2000 is configured to operate in a point-to-point (RS-232)
environment, the following will appear on the LCD when the RS-232
option is enabled:

COMMUNICATION MODE

1. Polled

2. Point to Point

3. Config Reports :2

he last line on the LCD indicates that timed reports may be utilized and
the present mode is point-to-point (:2). In this example, to reset the bath
to a polled environment, you would press:

[1] [ENTER]

Reports

The Reports option is accessed from the MODE submenu and is only
enabled when the Point-to-Point mode is selected.

Keypad Sequence (from initial LCD temperature display):

[MENU] [3] [ENTER] [2] [ENTER] [3] [ENTER]

The report is a serial dump of information on the operation of the CT-
2000 and includes:

„

instrument address

„

unit of temperature measurement (C/F)

„

a time reference

„

target bath temperature

„

actual bath temperature (as measured by the bath RTD)