Communication options – Cannon Instrument CT-2000 (Contant Temperature Bath) User Manual

Page 23

background image

21

CANNON

®

CT-2000 Constant Temperature Bath

Revision 1.0g—February, 2012; CANNON

®

Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

General Bath Offset

The General Bath Offset is a protected function of the CT-2000. This
option is a global adjustment which is made at the factory by
CANNON

®

. In rare instances when this setting may need to be altered,

the adjustment should be made in consultation with an authorized CAN-
NON

®

representative.

Communication options

Several communication options between the CT-2000 and your computer
may be configured from the MENU keypad option.

Change Port Speed

This option provides a speed setting for data transfer between the CT-
2000 and a computer via the RS-232C connection. Select the baud rate
(the default for most software is 9600) in bps (bits per second) from one
of the following options:

1. 1200

4.

19.2k

2. 2400

5.

38.4k

3. 9600

Keypad sequence (from initial LCD temperature display):

[MENU] [3] [ENTER] [1] [ENTER] [input speed choice (1-5)]
[ENTER]

Mode

The MODE keypad option permits RS-232C or RS-485 communication
options.

The CT-2000 is equipped with a full-duplex RS-232C point-to-point
interface as well as a half-duplex RS-485 network interface. Commands
and Queries may be sent to the CT-2000 via either interface. Responses
and Reports may be received by the master controlling computer.

Interface information

This interface is configured as 8 data bits with no parity and one stop bit.

The mode of operation, polled (multi-drop) or point-to-point, can only be
configured via the front panel menu.

Procedure

To select the MODE option, press 2 from the COM Setup menu.

KEYPAD SEQUENCE (from initial LCD temperature display):

[MENU] [3] [ENTER] [2] [ENTER]