beautypg.com

Adjusting the high temperature limit control, Custom temperature selection – Cannon Instrument CT-600 (Contant Temperature Bath) User Manual

Page 17

background image

15

CANNON

CANNON

CANNON

CANNON

CANNON

®

CT-500/600 Series II CONSTANT TEMPERATURE BATH

Version 3..2—March, 2007;

CANNON

CANNON

CANNON

CANNON

CANNON

®

Instrument Company

2139 High Tech Road • State College, PA • 16803 • USA

NOTE

If the auxiliary heater was engaged, the PREHEAT light will also blink.
Make sure to turn off the PREHEAT switch if operating at temperatures
below 80°C.

4. Insert the appropriate thermometer for the test temperature (see

APPENDIX C for thermometry information).

5. To adjust the bath temperature to the exact temperature desired, first

unlock the TA - FIXED dial.

6. Read the thermometer to determine the actual temperature of the

bath. Make sure that both the bath temperature and thermometer
have stabilized before taking a reading.

7. Turn the TA - FIXED dial to adjust the temperature. After each

adjustment, allow several minutes for the bath temperature to
stabilize.

8. Turn the dial clockwise to increase the bath temperature or turn the

dial counterclockwise to decrease the bath temperature.

9. When the desired temperature has been reached, re-lock the TA -

FIXED dial.

10. After the bath has reached the desired temperature, set the HIGH

TEMPERATURE LIMIT CONTROL (LC). This will prevent the
bath from overheating if a malfunction occurs.

Adjusting the High Temperature Limit Control

NOTE

During this procedure, the temperature of the bath may change slightly,
but will quickly recover.

1. Using a screwdriver, slowly turn the LC - TEMPERATURE

ADJUST control counterclockwise until the OVER TEMP message
lights up on the LC - PUSH TO RESET button.

2. Turn the control clockwise approximately ¼ to ½ turn.

3. Push the LC - PUSH TO RESET button. If the bath doesn't recover,

repeat step 2 and try again.

Custom temperature selection

To set the CT-500 to a temperature not indicated on the T-SELECT dial:

1. Turn the T - SELECT dial to the temperature closest to your desired

temperature. If the temperature is above the current bath tempera-
ture, the CT-500 will begin heating and the HEAT LED (Light-
Emitting Diode) will light.

2. Move the TA - SELECT switch to the right (the VARIABLE

position).

NOTE

To heat the bath more rapidly, turn on the PREHEAT (PH) switch. If the
target temperature is below 80°C, turn the PH switch off when the desired
temperature is reached. For temperatures above 80°C, keep the PH