7 .2 general inspection and cleaning instructions, Warning – BUCHI Recirculating Chiller F-125 User Manual
Page 23

7 Maintenance and repairs
23
F-125 Manual, Version B
7 .2
General inspection and cleaning instructions
Check the housing for visible defects (switches, plugs, enclosure etc.) and clean it regularly under
safe conditions with a damp cloth. Wipe off any splashes of aggressive chemicals immediately using
a damp cloth in order to avoid any damage being caused to the coating on the housing. Ethanol as a
cleaning agent is also possible to use.
Cleaning under safe conditions
Switch off the Recirculating Chiller and unplug the power cord.
•
Let the system acclimate to ambient temperature completely!
➡
Perform cleaning actions with a damp cloth.
•
Regularly clean lamella on front and rear side of the chiller with a damp cloth to remove dust, at
•
least once a year.
!
WARNING
Fire hazard, overheating, or triggering of the over-temperature protection function as a result of a
dirty air intake.
Clean the air intake as needed, but at least 1 x per year.
•
Hoses
Check the hoses for wear at least every six months.
•
Exchange damaged hoses.
•
Cooling media tank
Check prior to use the filling level of the Recirculating Chiller
•
Exchange the cooling liquid once a year by means of the drain valve and renew.
•
If you use a ethylene glycol / water mixture as cooling media you can drain it to the sink.
➡