beautypg.com

BUCHI Recirculating Chiller F-114 User Manual

Page 9

background image

2 Safety

9

F-100 / F-105 / F-108 / F-114 Operation Manual, Version E

Safety messages regarding Refrigerant R134a:

!

WARNING

Danger of injury and material damage due to overheating.
• Keep at a temperature not exceeding 45 °C

!

WARNING

Danger of corrosion and poisoning through inhalation of the fumes.
• In case of fire and/or explosion do not breathe fumes

!

CAUTION

If R134a escapes in the event of a fault..
• Avoid contact with skin and eyes
• Always wear safety goggles
• Always wear safety gloves

2 .5 .2

Personal protective equipment

Always wear personal protective equipment such as protective eye goggles and protective clothing.
The personal protective equipment must meet all requirements of all data sheets for the chemicals
used. These instructions are an important part of the Recirculating Chiller and must be made available
at all times to the operating personnel at the place where the equipment is deployed.

!

WARNING

Serious chemical burns by corrosives.
• Always wear protective goggles
• Always wear protective gloves
• Always wear laboratory coat

2 .5 .3

Built-in safety elements and measures

The instrument is fitted with a thermal overload protection for the compressor.
The presence of a minimum quantity of coolant is monitored by a fluid level sensor (F-108 and F114
only).