beautypg.com

3 .2 materials used – BUCHI Recirculating Chiller F-114 User Manual

Page 13

background image

3 Technical data

13

F-100 / F-105 / F-108 / F-114 Operation Manual, Version E

Technical data (100 V)

F-100

F-105

Power consumption (max.) 850 W

850 W

Supply voltage

100 VAC ±10 %

100 VAC ±10 %

Fuse

10 AT

10 AT

Frequency

50/60 Hz

50/60 Hz

Environmental conditions

For indoor use only

- Temperature

5-35 °C

5-35 °C

- Altitude

up to 2000 m

- Humidity
Display

No display

digital, resolution
0.1 °C

Overvoltage category

II

II

Degree of protection

IP20

IP20

Pollution degree

2

2

Instrument dimensions
W×H×D [mm]

280Ч590Ч420

280Ч590Ч420

Weight

37 kg

40 kg

Cooling power at 15 °C

400 W

500 W

Refrigerant

R134

R134

Temperature range

fix +10 °C

-5 °C ... +25 °C

Accuracy

±2 °C

±1 °C

Tank volume

3 L

3 L

Hose connection

8.0 mm

8.0 mm

Pump flow rate

2.5 L/min

2.5 L/min

Pump pressure

0.6 bar

0.6 bar

3 .2

Materials used

Materials used

Component

Material designation

Material code

Hazardous substances

Stainless steel, powder-coated with
polyester-epoxide

Housing

1.4301

Copper

Internal pipes and
coolers

CU

Polyester

Foil

PES

Glass-fiber-reinforced epoxy resin

Circuit board

Polyvinyl chloride

Cable

PVC

R134

Refrigerant

1,1,1,2-tetrafluoro-
ethane

Check MSDS