beautypg.com

I. introduction, Ii. specifications, Ii.1 power – Brookfield Temperature Controller, Model 75 User Manual

Page 3: Ii.2 controller, Ii.1 power ii.2 controller

background image

Brookfield Engineering Labs., Inc.

Page 3

Manual No. M/98-205-A0100

I. INTRODUCTION

The Model 75 Temperature Controller is part of the Brookfield Thermosel accessory to allow for
viscosity measurements at elevated temperatures. The unit includes a solid state proportional
temperature controller, a means for entering the desired temperature setpoint and appropriate
status indicators.

The front panel is used to display and enter setpoint temperatures. The front panel also presents
the status of the system (e.g. the current temperature of the Thermosel and any error messages).

A list of the parts shipped with the Model 75 Controller is provided below. If any parts are miss-
ing or damaged, please contact Brookfield or your local authorized representative immediately.

Model 75 System Contents

Description

Part Number

Quantity

Temperature Controller

HT-109

1

RTD Temperature Probe

DVP-94Y

1

Operating Instructions Manual

M/98-205

1

Recorder Output Cable

HT-88Y

1

A.C. Power Cord

DVP-65 (115v)

1

DVP-66 (230V)

1

TABLE 1

II. SPECIFICATIONS

II.1 Power

Voltage Specifications

85-265 VAC; 50/60 Hz

(check the label beneath the Temperature Controller
Input Voltage/Frequency for the voltage requirements of your unit)

TABLE 2

II.2 Controller

Temperature Controller Specifications

Range

above ambient to 300

°

C/572

°

F

Resolution

0.1

°

C or

°

F

Reading

±

1.0

°

C up to +150

°

C

Accuracy

±

2.0

°

C +150 to 300

°

C

Setpoint

The temperature will be maintained within 0.3

°

C of

Stability

the setpoint.

Recorder Output Voltage

0 to 4 Volts
1 volt = 0

°

C/32

°

F; 4 volts = 300

°

C/572

°

F

TABLE 3

Note:

The temperature accuracies stated above are a result of the combined accuracies of
the Model 75 Controller, the RTD temperature probe, and the Thermo Container.