Spindle Range Data
This Universal Spindle Range table lists the Spindle Range Coefficients for all (6) T-bar spindles. Dividing the coefficient number
by any rotational speed will give the full scale viscosity range for a Viscometer/Rheometer spindle/speed combination. (The Auto
Range key on DV-E, DV-I+, DV-II+ or DV=III+ instruments provides this information in the digital display).
Spindle Range Coefficient
Spindle
Entry Code
LV
RV
HA
HB
T-A
91
18,750
200,000
400,000
1,600,000
T-B
92
37,440
400,000
800,000
3,200,000
T-C
93
93,600
1,000,000
2,000,000
8,000,000
T-D
94
187,200
2,000,000
4,000,000
16,000,000
T-E
95
468,000
5,000,000
10,000,000
40,000,000
T-F
96
936,000
10,000,000
20,000,000
80,000,000
(Analog/Dial Viscometer)
Example: 1) Determine the full scale viscosity range (100% of scale) of a T-C spindle running on a RV Series @ 5 RPM.
Full Scale Range
=
Spindle Coefficient
=
1,000,000
= 200,000 cP
Spindle Speed
5 RPM
1) Determine minimum viscosity range (10% of full scale) at above conditions.
Min. Visc. Range
=
Full Scale Range
=
200,000
= 20,000 cP
10
10
Note: Maximum operable speed when using Helipath Stand is 10 or 12 RPM depending on speeds available on your
viscometer.
Viscometer Range Data
Range Data* (T-Bar Spindles) - applicable to Dial Reading & Digital Viscometers/Rheometers
DIAL READING
DV-I/DV-II
** DV-1+ / II+ /
DV-II+
Programmable
/ DV-II+ Pro
**DV-III / DV-III+ /
DV-III ULTRA
LV
156 - 3120
156 - 3120
156 - 9360K
156 - 9360K
RV
2K - 20K
2K - 20K
2K - 100M
2K - 100M
HA
4K - 40M
4K - 40M
4K - 200M
4K - 200M
HB
16K - 160M
16K - 160M
16K - 800M
16K - 800M
* Ranges in centiPoise (cP)
1 cP = 1 mPa•s
** Maximum range shown is at 0.1 RPM
K = 1000
M = 1,000,000