Spindle Range Data

This Universal Spindle Range table lists the Spindle Range Coefficients for all (6) T-bar spindles. Dividing the coefficient number
by any rotational speed will give the full scale viscosity range for a Viscometer/Rheometer spindle/speed combination. (The Auto
Range key on DV-E, DV-I+, DV-II+ or DV=III+ instruments provides this information in the digital display).

Spindle Range Coefficient

Spindle

Entry Code

LV

RV

HA

HB

T-A

91

18,750

200,000

400,000

1,600,000

T-B

92

37,440

400,000

800,000

3,200,000

T-C

93

93,600

1,000,000

2,000,000

8,000,000

T-D

94

187,200

2,000,000

4,000,000

16,000,000

T-E

95

468,000

5,000,000

10,000,000

40,000,000

T-F

96

936,000

10,000,000

20,000,000

80,000,000

(Analog/Dial Viscometer)

Example: 1) Determine the full scale viscosity range (100% of scale) of a T-C spindle running on a RV Series @ 5 RPM.

Full Scale Range

=

Spindle Coefficient

=

1,000,000

= 200,000 cP

Spindle Speed

5 RPM

1) Determine minimum viscosity range (10% of full scale) at above conditions.

Min. Visc. Range

=

Full Scale Range

=

200,000

= 20,000 cP

10

10

Note: Maximum operable speed when using Helipath Stand is 10 or 12 RPM depending on speeds available on your

viscometer.

Viscometer Range Data

Range Data* (T-Bar Spindles) - applicable to Dial Reading & Digital Viscometers/Rheometers

DIAL READING

DV-I/DV-II

** DV-1+ / II+ /

DV-II+

Programmable

/ DV-II+ Pro

**DV-III / DV-III+ /

DV-III ULTRA

LV

156 - 3120

156 - 3120

156 - 9360K

156 - 9360K

RV

2K - 20K

2K - 20K

2K - 100M

2K - 100M

HA

4K - 40M

4K - 40M

4K - 200M

4K - 200M

HB

16K - 160M

16K - 160M

16K - 800M

16K - 800M

* Ranges in centiPoise (cP)

1 cP = 1 mPa•s

** Maximum range shown is at 0.1 RPM
K = 1000
M = 1,000,000