Manual No. M/05-370-A1205
Position:
Range: 0-75 mm
Resolution: 0.1mm
Accuracy: 0.1mm
Output:
Computer: RS232 Compatible Serial Port
Environmental
0
°
C to 40
°
C temperature range (32
°
F to 104
°
F)
Conditions:
20% - 80% relative humidity, non-condensing atmosphere
I.5 Installation
1) Prepare a clean, level surface.
2) Unpack and remove the LFRA Texture Analyzer from the shipping container.
The LFRA Texture Analyzer weighs 15 kg (33 pounds). Use caution when lifting the
unit out of the packaging.
3) Place the LFRA Texture Analyzer on a sturdy, level surface. Adjust the four feet to ensure
that the instrument is stable.
4) Remove any additional components from the shipping package. Save the shipping container
and packaging for future use.
5) Make sure that the AC power switch at the rear of the LFRA Texture Analyzer is in the OFF
position. Connect the power cord to the socket on the back panel of the instrument and plug
it into the appropriate AC line. Position instrument so that the power cord can be removed
easily.
The AC input voltage and frequency must be within the appropriate range as shown
on the model and serial tag of the instrument (located on the back of the LFRA).
The LFRA Texture Analyzer must be earth grounded to ensure against electronic
failure!
6) If appropriate, connect interconnecting cable which is supplied with TexturePro Lite
software to the serial port for connection to a computer.
7) Turn the power switch to the ON position. The startup screen will indicate the firmware
version and load range of the LFRA (Figure II.1).
Figure I.1
8) Allow the instrument to warm up for 10 minutes. Check calibration according to Section
III.8.