I.2 utilities, I.3 units, I.4 specifications – Brookfield LFRA Texture Analyzer User Manual
Page 4: Units, Specifications
Brookfield Engineering Labs., Inc.
Page 4
Manual No. M/05-370-A1205
I.2 Utilities
Input Voltage:
115 VAC or 230 VAC
Input Frequency:
50/60 Hz
Power Consumption:
150 VA
Fuse:
Two 4 amp, 5 x 20mm, Time-lag
Power Cord Color Code:
United States
Outside United States
Hot (live)
Black
Brown
Neutral
White
Blue
Ground (earth)
Green
Green/Yellow
Main supply voltage fluctuations are not to exceed
±
10% of the nominal supply voltage.
I.3 Units
The LFRA Texture Analyzer uses the SI system of units for all parameters.
Parameter
Unit
Abbreviation
Load
Grams
g
Distance
Millimeter
mm
Time
Seconds
s
Speed
Millimeter per second
mm/s
I.4 Specifications
Load:
Range
Resolution
Model
Grams
Grams
Accuracy*
LFRA-100
0-100
0.01
±
0.5%FSR
LFRA-1000
0-1000
0.10
±
0.5%FSR
LFRA-1500
0-1500
0.20
±
0.5%FSR
LFRA-4500
0-4500
0.50
±
0.5%FSR
FSR = Full Scale Range
*This value applies to the complete range of operating temperatures as defined
in Section I.4. This value can be improved if the LFRA is controlled to a single
temperature.
Trigger Point:
Range
Resolution
Model
Grams
Grams
LFRA-100
0.1-10
0.01
LFRA-1000
0.2-100
0.10
LFRA-1500
0.2-150
0.10
LFRA-4500
0.6-500
0.10
Speed:
0.1 to 1.0 mm/s in increments of 0.1 mm/s
1 to 10 mm/s in increments of 1 mm/s
Accuracy:
±
0.1% of setpoint