beautypg.com

Brookfield KU-2 Viscometer User Manual

Page 6

background image

Brookfield Engineering Labs., Inc.

Page 6

Manual No. M04-242-D0612

I.3 Specifications

Range:

40–141 KU

32–1,099 gm

27 - 5,274 cP*

Accuracy:

± 1% of full scale range

Repeatability:

± 0.5% of full scale range

Paddle speed:

200 rpm ± 0.1 rpm

Printer output:

25D-Type Centronics Parallel Printer Output

Net weight:

22 lb. (10 kg)

Gross weight:

25 lb. (11 kg)

Dimensions:

15 x 11 x 20

Operating Environment: 0°C (32°F) to 40°C (104°F)

20% - 80% R.H.: non-condensing atmosphere

Electrical Certifications: Conforms to CE Standards for: Electromagnetic Compatibility

(EMC), Low Voltage (LVD) and Safety Requirements for electrical equipment for measurement

control and laboratory use.

*Centipoise scale is for reference only. Do not use for calibration.

I.4 Sample Container Specifications

Container Dimensions

Can Height

Bottom Flange Diameter

U.S. Pint

3.850" (9.78 cm)

3.385" (8.60 cm)

U.S. Half Pint

2.850" (7.24 cm)

2.875" (7.30 cm)

U.S. Quart

4.845" (12.31 cm)

4.230" (10.74 cm)

If you have a requirement for container dimensions other than those listed above, please contact

Brookfield or an authorized Brookfield agent.

I.5 Utilities

Input voltage:

115 VAC or 230 VAC*

Input frequency:

50/60 Hz

Power consumption: 16 VA (watts)

Power cord color code:

USA

Outside USA

Hot (live)

Black

Brown

Neutral

White

Blue

Ground (earth)

Green

Green/Yellow

*Main supply voltage fluctuations are not to exceed ± 10% of the nominal supply voltage.