beautypg.com

Crude fiber analysis, Important, Operator’s manual – ANKOM 2000 Series User Manual

Page 25

background image

Operator’s Manual


Rev E 1/14/15

pg. 25

Crude Fiber Analysis

Crude Fiber Calculation

Crude Fiber contained within a food or feed sample can be calculated using the following formula:

% Crude Fiber

=

100 x (W

3

– (W

1

x C

1

))

W

2

Where:

W

1

=

Bag tare weight

W

2

=

Sample weight

W

3

=

Weight of Organic Matter (loss of weight on ignition of bag and fiber)

C

1

=

Ash corrected blank bag factor (running average of loss of weight on
ignition of blank bag / original blank bag)

Crude Fiber Sample Preparation Procedure

To prepare samples for fiber analysis, follow the procedure detailed below.

IMPORTANT

When using the ANKOM

2000

Fiber Analyzer for Crude Fiber analysis, at least one blank

filter bag should be included with the sample set as an indicator of particle loss. A running
average of the blank bag weights is used in the fiber calculation as the C

1

correction factor.

A C

1

value larger than 1.0000 indicates that sample particles were lost from filter bags and

deposited on the blank bag(s). Any fiber particle loss from the filter bags will generate
erroneous results. If particle loss is observed, the grinding method should be evaluated.

1. Using a Solvent Resistant Marker, number all of the filter bags you will use

during the fiber analysis.

2. Weigh and record the weight of each empty filter bag (W

1

).

3. Set the Heat Sealer dial to between 4 and 5. (The setting may vary from sealer to

sealer.)

4. Seal at least one empty filter bag (to be used as a blank)

within 4mm of its open end. Keep the sealer arm down for 2
– 3 seconds after the red sealer light turns off (to cool the
seal). The seal can be seen as a solid melted stripe along the
top edge of the filter bag (as shown to the right). If the seal
is not strong, re-seal the bag.

Seal