Error reports, Technical reference manual, Disto – Leica Geosystems Leica DISTO OEM Module 3.0 WH15/WH30 User Manual
Page 8

Technical Reference Manual
V1.20
8/12
725538.doc
08.11.01
DISTO
DISTO
DISTO
DISTO
TM
TM
TM
TM
OEM module 3.0 WH15/WH30
OEM module 3.0 WH15/WH30
OEM module 3.0 WH15/WH30
OEM module 3.0 WH15/WH30
5.4. Error reports
No.
Format
Meaning
203
@E203
Prohibited parameter in command entry, or prohibited command, or non-valid
result
217
@E217
Parameter set-up incorrect (Contact Leica Geosystems)
221
@E221
Parity error (Prior to contacting Leica Geosystems please check the Terminal
settings)
222
@E222
Interface buffer overflow
(Contact Leica Geosystems if error occurs when sending less than 24
characters)
223
@E223
Interface framing error
(Contact Leica Geosystems)
224
@E224
GSI buffer overflow
(Contact Leica Geosystems if error occurs when sending less than 24
characters)
252
@E252
Temperature too high
(contact Leica Geosystems if error occurs at room temperature)
253
@E253
Temperature too low
(contact Leica Geosystems if error occurs at room temperature)
255
@E255
Received signal too weak, distance < 250mm
(Use different target and distances, if the problem persists, please contact Leica
Geosystems)
256
@E256
Received signal too strong
(Use different target and distances, if the problem persists, please contact Leica
Geosystems)
257
@E257
Too much background light
(Use different target and distances, if the problem persists, please contact Leica
Geosystems)
272
to
299
@E272
Hardware failure (Contact Leica Geosystems)
Before contacting Leica Geosystems please collect as much information as possible.