7 troubleshooting, Chapter 7 troubleshooting – Dillon Quick Balance Tension Meter User Manual
Page 24
24
Quick Balance User Instructions
7 Troubleshooting
Problem
Possible Cause
Solution
Powers on momentarily and turns off
Low battery
Replace with high quality alkaline batteries
Does not power on
Low battery
Replace with high quality alkaline batteries
Batteries installed backwards or no
spring contact
Insure that positive terminals of both batteries
(nub) face inward – towards the black cap.
Check that spring is attached to the battery
cap.
Software reset
Remove battery cap & reinstall after one
minute. Attempt to turn power on again.
Display contrast too light
Hold the Right Arrow key down while pressing
the F2 key several times to increase the
display contrast. If nothing occurs, release both
keys. Press the power button and try again.
Display is completely dark
Display contrast too dark
Hold the Arrow key down while pressing the F1
key several times to decrease the display
contrast.
Display drifts downward once installed
Wire material is creeping and internal
friction between wires is relieved.
This is normal behavior of wire. Lower display
resolution to mask this effect.
Temperature not accurate
Instrument changed temperature
environments
Allow instrument to remain in environment until
temp stabilizes or enter temp manually
Instrument exposed to sun
Enter temp manually