24

Quick Balance User Instructions

7 Troubleshooting

Problem

Possible Cause

Solution

Powers on momentarily and turns off

Low battery

Replace with high quality alkaline batteries

Does not power on

Low battery

Replace with high quality alkaline batteries

Batteries installed backwards or no
spring contact

Insure that positive terminals of both batteries
(nub) face inward – towards the black cap.
Check that spring is attached to the battery
cap.

Software reset

Remove battery cap & reinstall after one
minute. Attempt to turn power on again.

Display contrast too light

Hold the Right Arrow key down while pressing
the F2 key several times to increase the
display contrast. If nothing occurs, release both
keys. Press the power button and try again.

Display is completely dark

Display contrast too dark

Hold the Arrow key down while pressing the F1
key several times to decrease the display
contrast.

Display drifts downward once installed

Wire material is creeping and internal
friction between wires is relieved.

This is normal behavior of wire. Lower display
resolution to mask this effect.

Temperature not accurate

Instrument changed temperature
environments

Allow instrument to remain in environment until
temp stabilizes or enter temp manually

Instrument exposed to sun

Enter temp manually