4.1 TESTING FOR LEAKS
The Airtec System will provide maximum accuracy only if the system is leak free. Test
for leaks by applying a soap solution to system connections while the unit is under
normal operating pressure. If a leak is detected, tighten the fitting slightly and retest.
Repeat until the leak is stopped.
Warning
Do not over tighten connections. Return unit to factory if leaks persist.
Below is a list of potential leak sites that would have a direct effect on tube accuracy.
• Tube holder fitting
• Regulator/flow meter fitting
• Flow meter needle valve
• Flow meter cover tube
Leaks upstream (e.g. regulator fitting) although undesirable should not affect system
accuracy.
Note
Following leak checks allow 500 cc/min to flow through the flow meter for 10 minutes to
remove any moisture accumulation.
Note
Leak checks should be run 500 cc/min with a used tube in place (use tube #109AD).
4.2 EQUIPMENT FAILURE
In the event of failure of the regulator or flow meter return the unit to Nextteq LLC for
servicing. Do not attempt to repair either the regulator or the flow meter. For
information on equipment returns, see Appendix B: Returned Material Authorization
(page 24).