Nextteq llc (rev. 2) 4/09 – Nextteq Standard Airtec Regulator Kit User Manual
Page 13
![background image](/manuals/646802/13/background.png)
SECTION I: Introduction
13
NEXTTEQ LLC (Rev. 2) 4/09
1.4
ACCURACY
&
CALIBRATION
The Airtec System when used in accordance with tube instructions provides an accuracy
of
±25%.
The flow meter is factory calibrated and should not require recalibration with normal use
and proper care of the unit.
Should a calibration check be required, the unit may be returned to the factory, or
calibration may be checked using a soap film bubble meter.
TABLE 2
AIRTEC TUBE DATA
Gas/
Substance
Part
No.
Measuring
Range
Flow
Rate
(cc/min)
Sample
Time
(Min)
Humidity
Range
(%RH)
Color Change
Carbon Monoxide
1A
5-50 ppm
100
3
20-90
Yellow to Dark Brown
Carbon Dioxide
2AG
200-3000 ppm
100
1.5
Pale Blue to Purple
Water Vapor
6AH
500-5000 ppm
300
1
Green to Purple
Water Vapor
6A
30-80 mg/m
3
100
10
--
Yellow to Purple
Water Vapor
6AG
150-3000 mg/m
3
300
1
--
Green to Purple
Nitrogen Oxides
11A
0.02-0.7
0.06-2
100
100
5
2
White to Bluish Green
Oil Mist *
109AD
0.2-5.0 mg/m
3
1000
20
Orange to Pale Blue
* Oil Mist test may be run 1000 scc/min for 20 min as indicated in tube instructions.
Flow rates over 500 scc/min require a high range flow meter. However, Nextteq’s
Regulator has a maximum flow rate of 500 scc/min, so the oil mist tubes should run at a
500 scc/min for 40 minutes.
All tubes have a Temperature Range of 4-40
°C.