Error or Indicator
Cause
Explanation / Solution
- - - -
Below zero
Re-zero the scale by pressing the
Ж0Е key.
- -
Taring is not allowed
Place a weight on the pan before
taking a tare.
- -
Re-zeroing is not allowed
Remove weight from pan and clear
stored tare values.
L
Pan error
Make sure that pan is properly
seated on the pan support.
H
Exceeded capacity
The scale has exceeded its
weighing capacity. Reduce the
weight.
Err - b
Pan not cleared on power-up
The pan was not empty while the
scale was initiating at startup.
Clear the pan and restart the scale.
Sapl LO (Parts Counting)
Average piece weight is too low
The average piece weight is lower
than 3e. Be sure that individual
piece weight is greater than 3e.
Sapl LO (Percent Weighing)
Sample reference weight is too low The sample reference weight is
lower than 100d. Increase the
sample weight.
The scale will not turn on
Possible power failure
Check the power AC adapter
connection.
Weighing results are not accurate Inaccurate calibration
Calibrate the scale.
Weighing result will not stabilize Irregular environment
Eliminate drafts or vibrations.
Cleaning and maintaining your Scale:
• Do not use chemicals or benzene when cleaning the surface. Corrosive chemicals may damage the
finish.
• Alcohol may be used only to clean the scale’s stainless steel surfaces.
Chapter 29: Maintenance
Chapter 28: Common Errors and Troubleshooting