8.2. Calculation Formula
1. Use the “<>” navigation key to select Calculation and press the ENTER key.
2. The current moisture analysis formula will be displayed. Use the “<>” navigation key to alternate
between formulas. Once a desired formula is displayed, press the ENTER key to confirm the selection.
Where m
0
= initial (wet) weight, m = current weight, and W [%] = moisture as a %:
1. W [%] = m
0
– m/m
0
* 100%, where moisture is determined in relation to initial weight
2. W {%} = m
– m/m * 100%, where moisture is determined in relation to current weight
3. W [%] = m/m
0
* 100%, where current weight is determined as a percentage of sample weight
8.3. Drying Temperature
1. Use the “<>” navigation keys to select Drying Temp and press the ENTER key.
2. Use the Numeric keys to enter the desired Drying Temperature in degrees Celsius. The maximum
temperature is 160°C.