Perc – Torbal BTA Series User Manual
Page 25
- 24 -
Function Options:
PER Off
( Peroff)
Disables Percent Weighing
PER On
(per on)
Enables Percent Weighing
Percent weighing is used to express the weight of an object as a percentage of a stored sample weight. To use
percent weighing, follow these steps:
1. Place container on the pan and press the T key to tare.
2. Once the scale has been tared, place the sample reference weight on the pan.
3. The weight of the sample will be displayed as it is placed on the pan. Once the weight of the
sample stabilizes and the stability indicator appears on the display, press the MENU key to
enter the main menu.
4. When “PERCENT” is displayed, press the T key.
Chapter 13: Percent Weighing (PerC)
-------
15.23
PerC