“Calibration Required

ONLY

if Rear- LASER Adjustment Set Screw has

been tampered with”


Tools Required: Elevated Surface (wheel alignment, frame machine, etc.)

Level (carpenters level or other means)

6‘ Straight Edge (board or other means)

5/64” Allen Wrench

Locate an elevated surface and lay the straight edge on surface facing away from
optical head. Check straight edge for level, shim if necessary.

Move SCA 1

ISOColor to end of straight edge, turn on the laser and adjust

height of optical head so that the mounted laser will shoot down the straight edge.









Adjust the rear floor slope wheel until the level in the optical head is centered. Readjustment of
height of optical block may be necessary.

Adjust the rear height adjustment set screw of the laser assembly (Lock Tight has been installed
on screw at factory, minor pressure should break seal) till laser is viewed at both ends of straight
edge equally.

Laser is now calibrated, installing lock tight or other adhesive to adjustment screw is
recommended.

5.2 MAINTENANCE


The SCA 1

ISOColor will provide years of trouble free operation with minimum maintenance, however,

care should be taken in the day-to-day usage of this service instrument. Following are areas that should be
periodically checked and serviced;

Check wheel axles nuts for tightness, minor lubrication is
recommended.

Check the optical head mounting bolts for tightness and tighten as
required.

Check all other mounting screws, bolts and nuts for tightness.

Clean the mast area where the brake rides with a mild detergent to
assure of secure holding.

Clean the front lens, sighting unit and viewing window with a mild
detergent being careful to use a non-abrasive soft cloth.

Visually inspect alignment screen for burn holes that could have been
caused by exposure to direct sunlight. Caution should always be
exercised in storing of SCA 1, a placement of a dust cover (P/N:
01016000) on unit is recommended during storage.


6. FREQUENTLY ASKED QUESTIONS


Question: Level in optical head is not centered during alignment procedure?

Answer:

Level vial is used ONLY when checking the calibration of the FLOOR SLOPE LASER.

Question: The high intensity (hot spot) area of one headlamp is of a different color than

the other lamp?