Isotropic Electric Field Probes
HI-4453/FP5083/FP4083/FP2083
© ETS-LINDGREN L.P. – June 2005
Rev. F, P# H-600064
31
7. APPENDIX B: HI-
4453/FP5083/FP4083/FP2083
OPERATING PROTOCOLS
The information in this appendix assumes that you have
purchased the optional HI-4413P Fiber Optic/RS232 Interface,
and are capable of communicating directly with the HI-
4453/FP5083/FP4083/FP2083 probe via computer. No system
readout is required when using this configuration.
COMMUNICATION PROTOCOL
Data Type:
RS-232 Serial
Data Mode:
Asynchronous
Word Length:
7 bit
Parity:
Odd
Stop Bits:
1
Data Rate:
9600 baud
INFORMATION TRANSFER PROTOCOL
The HI-4453/FP5083/FP4083/FP2083 operates as a Controller
Mode device. This probe only responds to commands from
another device; it transmits no data without receiving instructions
to do so.
Command Structure
A command to an HI-4453/FP5083/FP4083/FP2083 probe
consists of
1) A command letter, followed by
2) Possible
parameters
3) Terminated with a carriage return.