beautypg.com

ETS-Lindgren 7405 E & H Near Field Probe Set User Manual

Page 38

background image

38

Common Diagnostic Techniques

www.ets-lindgren.com

Differential mode data is generally well behaved. The amplitude

measured with the H-field probe will be significantly higher than that

measurement with the E-field probe. Also, the H-field will drop off at a

much faster rate than the E-field.

Common mode measurements are generally less well behaved. Often

the best indicator is the relative amplitude. The E-field probe will have

a much higher reading than the H-field probe. The drop-off rate will be

faster when measured with the E-field probe. However, experience

shows that the E-field, being a high potential field, is much more

susceptible to perturbation. Often the reading will be sensitive to cable

placement and differences in the position of the person holding the

probe. This susceptibility to being perturbed can be a hint that the field

is coming from a high potential source.