Spectrum Analyzer / Receiver

Use the BNC connectors on the front panel to connect the spectrum

analyzer or EMI receiver.

The monitor port which is not being monitored should be terminated

into a coaxial 50-ohm terminator.

Connect the input and output terminals to their proper mains

connections and EUT connections before connecting the monitor port

to the measurement instrumentation; otherwise, power surges may

damage the test instrumentation mixers or attenuators.

Removal of a terminator and connection to the BNC receptacle will not

generate power surges.

When the power is to be disconnected, remove the coaxial connection

to the measurement instrumentation first to avoid possible damage.