The Model 4825/2 is provided with resistors to

help bleed off high voltage transients, but it is

advisable to connect the input and output

connectors to their proper power lines and

loads before connecting the monitor port to the

measurement instrumentation; otherwise,

power surges or transients can damage the test

instrumentation mixers or attenuators.

The Model 4825/2 Line Impedance Stabilization Network (LISN) is designed for

use in Installation Category II and Pollution Degree II per lEC-1010 and IEC-664.

When installing in a cabinet, make sure that the convection around the product is

not restricted. The ambient temperature outside the cabinet must be less than

the maximum operating temperature of the Model 4825/2 by 4

°

C for every

100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is

greater than 800 watts, then forced convection must be used.

Model 4825/2 Front View

Model 4825/2 Back View