background image

4.0 Installation

Before connecting any components, follow the

safety information in the ETS-Lindgren

Product Information Bulletin included with your

shipment.

Overcurrent protection is provided in the

Model 4825/2. The rating of the integral 2-pole

circuit breaker is 250 VAC 50/60 Hz at

25 amperes. If overcurrent protection at a lower

rating is required by the Equipment Under Test,

the unit must be connected to a power mains

which has appropriately rated mains protection

installed.

The Model 4825/2 is provided with a protective

earthing connection integral to the power inlet

and input power cord assembly. The mains plug

should only be connected to a mains source

which utilizes a protective earth conductor. Due

to the high leakage current to ground inherent

in this type of equipment, it is necessary to

install a supplemental protective earthing wire

from the protective earth terminal on the rear

panel to an appropriate earthing point on the

power mains. This earthing point should be

determined by an electrician authorized to

perform such work by appropriate code or law.

Any interruption of the protective conductor

inside or outside of the unit is likely to make the

Model 4825/2 dangerous. Intentional

interruption is prohibited. The supplemental

ground wire is supplied with the unit.

Installation