Line select switch – ETS-Lindgren 3816-2 LISN User Manual

Page 15

background image

www.ets-lindgren.com

Installation and Application

15

L

INE

S

ELECT

S

WITCH

Select the line to be monitored by the three-position selector switch. The line not

selected is internally terminated into 50 ohms. When using the remote (center)

position, connect the monitored line to the 9-pin connector on the rear panel.

Control signals are low-level 5 volt DC.

To monitor Line L1, apply +5VDC to pin 1 of the 9-pin connector.

To monitor L2, apply +5VDC to pin 3.

Should +5VDC be applied to both pins, the first pin to receive the

control signal will be selected.

As with manual selection, the line not selected is terminated into 50

Ω.

Pin 2 of the 9-pin connector is control common.

Although the unit is provided with resistors to

help bleed off high voltage transients, connect

the input and output connectors to their proper

power lines and loads before connecting the

monitor port to the measurement instrument.

Otherwise, you may damage the mixers or

attenuators of the test instrumentation due to

power surges or transients.

Switching between the two lines will not generate transients. Remove the power

source first when disconnecting power.