5.0 Data

Graphs of the calibration data for each measurement port of the Model 3810/2

Line Impedance Stabilization Network (LISN) are included with the unit. The

graphs provide individual plots of both impedance and insertion loss data.

Impedance is plotted in a semi-log format where frequency is displayed on the

horizontal from 9 kHz to 30 MHz. The vertical has a range of 0 ohms to

100 ohms and represents the measured impedance of the unit. The insertion loss

is also plotted with frequency on the horizontal from 9 kHz to 30 MHz. The

vertical of the graph has a range of -9 dB to 1 dB and represents the measured

insertion loss of the Model 3810/2.

A Certificate of Calibration Conformance is provided with each Model 3810/2.

Data

| 19