Earth line choke switch, Ac receptacle, Artificial hand – ETS-Lindgren 3810-2 LISN User Manual

Page 16

background image

E

ARTH

L

INE

C

HOKE

S

WITCH

The safety ground isolation choke selector switch switches the 1.6 mH earth line

choke IN and OUT of the safety ground circuit. The ground choke is designed

and manufactured with sufficient capacity to conduct the maximum current rating

of the Model 3810/2 and at no time is the safety ground of the unit compromised.

The earth line choke avoids a double RF ground connection (safety ground and

measurement ground) in the conducted emissions test setup.

AC

R

ECEPTACLE

Connect the Equipment Under Test (EUT) to the panel-mounted AC receptacle.

The style of receptacle is determined by the model specified. Following are the

standard output receptacle types:

NEMA (Type 5-15R)

Schuko (Type CEE 7/7)

British Standard (BS 1363)

Australian (AS 3112)

A

RTIFICIAL

H

AND

In conformance with EN55014 and BS800, the artificial hand connection is used

to test handheld equipment that is provided without earth connections.