Programming auto clear – CDI Torque MULTITEST Motorized Loader User Manual
Page 34

3-7
Setup and Programming
Setting High Torque/Force Limit
1. Press SET HIGH.
—
The LED display reads:
last SET HIGH value , with one digit flashing
—
The LCD display reads:
Edit high limit - upper line
[units] - lower line
Change units of measure by pushing UNITS.
2. Select an LED digit to change (flashing) by pushing left
or right DIGIT.
3. Increment or decrement the selected digit by pushing up
or down VALUE. Push and hold up or down VALUE to
change the display quickly.
A decimal point "." appears between "9" and "0"
numerals. Maximum resolution is selected
automatically regardless of decimal point position
upon pushing ENTER. The program value is
absolute, for both ± readings.
4. Select and change the next LED digit using left and right
DIGIT and up and down VALUE. Continue until all digits
are set.
5. Press ENTER to return to measurements mode.
Programming AUTO CLEAR
Use AUTO CLEAR to program the CLEAR, STORE and SEND
functions for automatic or manual operation. Use these
functions in PEAK HOLD, POWER TOOL and FIRST PEAK
modes to STORE torque or force measurements to memory,
CLEAR the display, and SEND the last measurement captured
on the LED display to the printer or PC. Auto STORE and
SEND are initiated at either AUTO or delayed CLEAR
operations.
Use the manual modes to:
• Store the present reading by pushing STORE DATA,
• Clear a captured display by pushing MANUAL CLEAR, or
• Send a measurement to the printer/port by pushing
SEND/AUTO SEND.
Use the automatic modes to:
• Store the last measurement, and/or
• Send the last measurement to the printer/port by initiating the
auto or delayed CLEAR function.