Salter Brecknell 6712 User Manual
Page 15

Model 6712 Point-of-Sale Interface Scale User’s Manual
15
5. Press the ZERO key to scroll through the
available capacity selections. See Table 1 for
available capacity selections. When the
choice you want is displayed, press the
TEST key. . .
LOAD 0 is displayed.
lb
kg
oz
5 x 0.005
2.5 x 0.002
80 x 0.05
15 x 0.01
7 x 0.005
160 x 0.1
30 x 0.02
15 x 0.01
800 x 0.5
50 x 0.05
25 x 0.02
1600 x 1
100 x 0.1
50 x 0.05
150 x 0.05
75 x 0.02
150 x 0.1
75 x 0.05
6. Clear all weight from the weigh platter and
press the TEST key. After a brief wait…
LOAD 10 is displayed (or the current
capacity selected in Step 5). Alternate
calibration points can be chosen using
the ZERO key to scroll choices.
The NCI 6712 allows calibration using less
than full capacity weights. See Table 2 for a
list of the alternative weights that can be
used to calibrate your scale for its designed
capacity.
Make sure the weigh
platter and access
plate are on the scale
to establish proper
zero load.
Table 1
Capacity Selection