TROUBLESHOOTING CHART

Symptom

Possible Cause

Corrective Action

Container is cold and 1. Vacuum loss. Check NER.

1. Consult with Taylor-Wharton

may have ice or frost

for course of action. Do not

on outer casing. Will

attempt to put additional

not hold liquid over-

liquid in container.

night. Relief valve is

2. Defective P.B./Economizer

2. Look for P.B. coil pattern in

venting gas.

regulator.

ice. Close P.B. valve.
Replace or reset regulator.

Ice formation on

1. Pressure building valve

1. Replace or rebuild valve.

bottom of container

not closing properly.

when P.B. valve is

2. Leak in pressure

2. Leak test piping

closed.

building system

connections and tighten

topworks.

fittings if needed.

Container vents

Pressure Building/

Remove and reset or replace

through relief valve

Economizer Regulator set

regulator.

when in use.

above relief valve setting.
Economizer side of
regulator clogged or
stuck open.

Container vents after

This may be caused by

Symptom should go away

fill but quits after

residual heat vaporizing

once container reaches

awhile.

some liquid inside

operating temperature and

container and is a normal

the liquid reaches its

condition.

saturation point at container
operating pressure.

Container vents gas

Heat leak may be too

Perform container

continuously through

great.

performance evaluation test

relief valve.

per Maintenance section to
determine if container
vacuum is adequate.

Level indicator stuck

Float rod stuck on or in

Reinstall. See Contents

1/2 full. Yellow

float rod guide.

Gauge Installation.

indicator ring will not
move.

Level indicator at

Indicator disengaged from

Recouple indicator using

bottom of gauge.

gauge rod. Caused by

engagement ring.

Container full of

dropping the container.

product.