beautypg.com

Moving the tank – Taylor-Wharton Bulk Tank Safety Data User Manual

Page 15

background image

15

Casing Safety Device

When the safety device has ruptured or leakage is caused by corrosion or

other damage, the safety device assembly must be replaced

Warning

Do not start any cutting of welding work until gaseous concentration in the

insulation space has been checked for combustibles. If necessary, purge

insulation space with dry nitrogen gas until combustible gases have been

eliminated. Do not allow pressure in the insulation space to exceed the

pressure in the inner container. External overpressure on the inner con-

tainer could cause permanent damage.

Because re-evacuation is time consuming and usually

requires taking the tank out of service, it is not normally

attempted until tank performance becomes unacceptable.

Even a relatively high degree of deterioration can be

tolerated in a tank from which high rates of withdrawal

are being made. However, if vacuum deterioration

seriously affects tank operation by producing excessive

pressure buildup and high loss rates, re-evacuation may

be warranted.

Note

Do not use Teflon tape as a sealant on vacuum system fittings

When moving a tank from another location after being in cryogenic service,

be sure that the tank is empty and warm. If the tank has been in oxygen or

hydrogen service, purge it warm filtered nitrogen gas to eliminate any

residual combustible gas. To prevent contamination from atmospheric

moisture, pressurize the tank to approximately 20 psig (138 kPa/1.4 bar)

with filtered nitrogen gas and close all valves to retain pressure.

Vertical tanks should be shipped in the horizontal position on wooden

cradles. Shipping drawings may be purchased from the Theodore, Alabama
factory. Give tank model and part numbers when requesting drawings.

Moving the Tank