Trial run – Scilogex MS7-H550-Pro User Manual
Page 9

7
Digital 7” Square(Hotplate)Magnetic Stirrer
Remote
Display in case of remote control.
Err
Display in case of error happening.
Rated value/Real
value
Display value in case of heating or
stirring function switching ON.
Table 3
Note:
If both heating and stirring functions have been
started at the same time, heating function
always has higher priority. If in this case speed
is changed via the stirring knob, it displays
stirring speed and reverses to temperature in
the duration of 5 seconds.
5.Trial run
•
Make sure the required operating voltage and power
supply voltage match.
•
Ensure the socket must be properly grounded.
•
Ensure the power is off.
•
Plug in the power cable,ensure the power is on and begin
initializing.
•
Add the medium into the vessel with a stirring bar.
•
Place vessel on the work plate.
•
Set the rated stirring speed and begin.
•
Observe the stirring bar and LCD display.
•
Set the rated temperature and start heating.
•
Observe the real temperature on LCD display.
•
Stop the heating and stirring functions.
If these operations above are normal, the device is ready
to operate. If these operations are not normal, the device
may be damaged during transportation, please contact