Faults – Scilogex MS7-H550-Pro User Manual
Page 13

11
Digital 7” Square(Hotplate)Magnetic Stirrer
The unit can be controlled from an external PC (using the
dedicated software) via the RS232 C serial interface fitted to
the unit. Data communication from laboratory instrument to
computer is only possible on demand of the computer.
•
The functions of the interface lines between laboratory
instrument and automation system are selected from
the specified signals of the EIA-standard RS232C,
corresponding with DIN66020 Part 1. The allotment of
the bushing can be taken from Figure 7.
•
Transmission method: Asynchronous signal transmission
in start-stop-operation.
•
Mode of transmission:Fully Duplex.1start bit; 7 character
bits; 1stop bit.
•
Transmission speed: 9600 bit/s.
•
Start remote control knob LED display “Remote”.
Note:
Forbid to insert or remove the RS232C
communication line when switch on!
10.Faults
•
Instruments can not be power ON
-
Check whether the power cable is plugged in
-
Check whether the fuse is broken or loose
•
Fault in power on self test
-
Switch OFF the unit, then switch ON and reset the
instruments to factory default setting.
•
Temperature can switched on not reach set point
-
Check whether the safety temperature value is set too
low.
•
Stir speed can not reach set point
-
Check whether the stirring bar breakaway monitoring
is ON.
-
Excessive medium viscosity may cause abnormal
speed reduction of the motor.
•
Heating or stirring function do not start when knob is
pressed.
-
Check whether the unit is in the mode setting program,
-
Exit from the mode setting and restart the heating/
stirring functions.
•
Unit do not power down when switched off.
-
Check if the residual heat warning function is still ON
and hotplate temperature is above 50℃(the LCD still
works and“Hot”flashes).
If these faults are not resolved,please set the instruments
to factory default setting or take the unit to technical service
center.