MS-(H)-Pro Digital(Hotplate)Magnetic Stirrer

5.2 Display

Temp

Stir

Remote

Set

Probe

Hot

C

B

A

Err

Real value

Rated

value

Temperature Stir

Error

Set

Mode

HotPlatewar

ning

External

probe

Remote

control

Figure 4

Characters

Descriptions

Temp and ℃

Display temperature when the heating function is
switched ON.

Stir

Display stirring state when the stirring function is
switched ON.

Mode

Display the setting mode in order of A, B and C
after pressing mode key.

Set

Press setting key to set function, and display Set characters.

Hot

Display hot warning if the heating plate temperature is
above 50℃ after switching OFF the heating function.

Probe

Display when using external probe.

Remote

Display in case of remote control.

Err

Display in case of error happening.

Rated value 、
Real value

Display value in case of heating or stirring function
switching ON.

Table 3

Note:

If both heating and stirring functions have been

started at the same time, the display of function

heating” always has higher priority. If in this

case speed is changed via the stirring knob, it

display stirring speed and reverses to temperature

in the duration of 5 seconds.

6 Initialization

6.1 Initializing sequence

• Turn on instrument by the power switch.

• The “Heat” and “Stir” LEDs and the LCD display

flash about 3 times.

• LCD displays the workplate

,

s safe temperature limit in

SAFE xxx°C”.

• LCD displays residual heat warning “rES On/OFF”.

• LCD displays stirring bar breakaway monitoring “br On/

OFF”.

• LCD displays operating modes(A、B、C).

• LCD displays the “Probe” characters in case of

connecting with external temperature sensors.

6.2 Default settings

MS-(H)-Pro is supplied with the following default settings

This manual is related to the following products: