6 initialization, 2 display, 1 initializing sequence – Scilogex MS-Pro User Manual
Page 9: 2 default settings

7
MS-(H)-Pro Digital(Hotplate)Magnetic Stirrer
5.2 Display
Temp
Stir
℃
Remote
Set
Probe
Hot
C
B
A
Err
Real value
Rated
value
Temperature Stir
Error
Set
Mode
HotPlatewar
ning
External
probe
Remote
control
Figure 4
Characters
Descriptions
Temp and ℃
Display temperature when the heating function is
switched ON.
Stir
Display stirring state when the stirring function is
switched ON.
Mode
Display the setting mode in order of A, B and C
after pressing mode key.
Set
Press setting key to set function, and display Set characters.
Hot
Display hot warning if the heating plate temperature is
above 50℃ after switching OFF the heating function.
Probe
Display when using external probe.
Remote
Display in case of remote control.
Err
Display in case of error happening.
Rated value 、
Real value
Display value in case of heating or stirring function
switching ON.
Table 3
Note:
If both heating and stirring functions have been
started at the same time, the display of function
“
heating” always has higher priority. If in this
case speed is changed via the stirring knob, it
display stirring speed and reverses to temperature
in the duration of 5 seconds.
6 Initialization
6.1 Initializing sequence
• Turn on instrument by the power switch.
• The “Heat” and “Stir” LEDs and the LCD display
flash about 3 times.
• LCD displays the workplate
,
s safe temperature limit in
“
SAFE xxx°C”.
• LCD displays residual heat warning “rES On/OFF”.
• LCD displays stirring bar breakaway monitoring “br On/
OFF”.
• LCD displays operating modes(A、B、C).
• LCD displays the “Probe” characters in case of
connecting with external temperature sensors.
6.2 Default settings
MS-(H)-Pro is supplied with the following default settings