Drucker Diagnostics Model 755 Centrifuge User Manual
Page 6

6
Specifications: 12-Place
Speed Range (Model 755V):
755V w/12-Place Horizontal Rotor:
500 to 3,200 (± 100) RPM*
755V w/Fixed-Angle Rotors:
500 to 4,000 (± 100) RPM*
Max Speed (Model 755): (This speed is set by the factory, see page 7)
with 12-Place Horizontal Rotor
3,600 (± 100) RPM
with Fixed-Angle Rotors
3,600 (± 100) RPM
Maximum capacity:
12-Place Horizontal Rotor:
180 mL (12 x 15 mL)
20-Place Fixed-Angle Rotor:
300 mL (20 x 15 mL)
Overall Dimensions (H x W x D):
9 in. x 14.5 in. x 17 in.
Centrifuge Motor:
1/2 H.P. Brushless DC
Nominal Acceleration Time:
30 seconds
Nominal Braking Time:
15 seconds
Protection Breaker:
4 Amp. re-settable
Timer (electronic):
1 to 30 min ± 1%
Current Requirement:
2.0 A @ 115 VAC
or 1.0 A @ 230 VAC
Voltage Requirement:
115 or 230 (±10%) Volts AC
Frequency:
50 / 60 Hz
Weight:
39 lbs.
Specifications: 24-Place
Speed Range (V.S.V.):
500 to 3,150 (± 5%) RPM
755V w/24-Place Horizontal Rotor:
3,150 (± 100) RPM*
755V w/20-Place Fixed-Angle Rotors:
3,200 (± 100) RPM*
Force Range (variable speed version):
45 to 1,760 (± 10%) xg
Single Speed Setting:
(This speed is set by the factory, see page 7)
3,000 (± 5%) RPM
Force (@ 3000RPM):
1,600 x g
Maximum capacity:
240 mL (24 x 10 mL)*
300 mL (6 x 50 mL)*
Overall Dimensions (H x W x D):
9 in. x 14.5 in. x 17 in.
Centrifuge Motor:
1/2 H.P. Brushless DC
Nominal Acceleration Time:
30 seconds
Nominal Braking Time:
15 seconds
Protection Breaker:
4 Amp. re-settable
Timer (electronic):
1 to 30 min ± 1%
Current Requirement:
2.0 @ 115 VAC
or 1.0 A @ 230 VAC
Voltage Requirement:
115 or 230 (±10%) Volts AC
Frequency:
50 / 60 Hz
Weight:
39 lbs.
* Any use other than those specified by the manufacturer is explicitly prohibited.
Maximum speed shown is for the given rotor with 100mm red tube holders. The use of other tube holders
will alter the maximum speed.