STERILIZATION INSTRUCTIONS
• Miltex recommends that the sterilization trays be processed according to the sterilization wrap
manufacturer’s instructions prior to sterilization to maintain sterility of internal components/items
and for proper aseptic presentation to the surgical/dental field.
• For Prevacuum Sterilizer:
o
Wrapped trays and instruments should be exposed to 132° C to 135° C (270° F to 275° F)
for at least four minutes.
o
Dry 20 to 40 minutes.
• For Gravity Displacement Sterilizer
o
Wrapped trays and instruments should be exposed to 132 C to 135 C (270 F to 275 F) for at
least 30 minutes, or 121° C to 123° C (250° F to 254° F) for at least 55 minutes.
o
Dry for 20 to 50 minutes.
• Variables that may affect drying times include: loading density of the tray, instrument configuration,
total contents of the sterilizer, steam quality, equipment maintenance and others.
MAINTENANCE OF TRAYS
• Always inspect for cleanliness or damage before use.
• Make sure all latches and handles are secure and in working order.
SYMBOLS USED ON LABELING
Manufacturer
Authorized Representative in the European Community
Catalog number
Lot number
Non-sterile: Sterilize prior to use
CAUTION: See warnings or precautions
Consult instructions for use
Product complies with requirements of directive 93/42/EEC for European Union Class 1
medical devices
CAUTION: Federal (USA) law restricts this device to sale by or on the order of a
dentist.
Miltex, Inc.
Miltex GmbH
589 Davies Drive
78604 Rietheim-Weilheim
York, PA 17402
Germany
800-645-8000
©2008 Miltex, Inc. All rights reserved
.
www.miltex.com
SH-030000 Rev.1
10/08