Cooper Instruments & Systems Series M3 Digital Force Gage User Manual
Page 14
Series 3 Digital Force Gauges
User’s Guide
13
3. To escape the Calibration menu at any time, press ESCAPE. The display will appear as follows:
Selecting “CANCEL” will revert back to the Calibration setup. Selecting “EXIT W/O SAVING” will
return to the menu without saving changes.
4. After the number of calibration points has been entered, press ENTER. The display will appear as
follows:
5. Place the force gauge horizontally on a level surface free from vibration, then press ZERO. The
gauge will calculate offsets, and the display will appear as follows:
If failed:
6. The following screen appears after the offsets have been calculated:
Attach weight fixtures (brackets, hooks, etc), as required. Do not yet attach any weights or apply
any calibration loads. Then press ENTER.
CALIBRATION
NOT COMPLETE
CANCEL
EXIT W/O SAVING
CALIBRATION
COMPRESSION
Attach necessary
weight fixtures.
THEN PRESS ENTER
CALIBRATION
OFFSET
Sen.Offset Adj.Failed
Ana.Offset Adj.Failed
CALIBRATION
OFFSET
Sen.Offset Adj.Passed
Ana.Offset Adj.Passed
CALIBRATION
OFFSET
Please wait…
CALIBRATION
OFFSET
Place force gauge
horizontal
THEN PRESS ZERO