Cooper Instruments & Systems TT03C Digital Cap Torque Gage User Manual
Page 16

Series TT03 Digital Torque Gauges
User’s Guide
15
Remove the torque applied in Step 7, leave the fixtures in place, then press ZERO.
9. The display will appear as follows:
Use the UP and DOWN keys to adjust the torque value as required. The torque values default to
even increments, as indicated by the previously entered number of data points described in Step
1. Then press ENTER.
Repeat the above step for the number of data points selected.
10. After all the clockwise calibration points have been completed, the display will appear as follows:
Press ENTER.
11. At the completion of the counter-clockwise calibration, the display will appear as follows:
To save the calibration information, select “SAVE & EXIT”. To exit without saving the data select
“EXIT W/O SAVING”.
12. Any errors are reported by the following screens:
Displayed at the start of calibration if a disallowed unit is selected.
CALIBRATION
Units must be kgFmm.
PLEASE TRY AGAIN
PRESS ENTER
CALIBRATION
COMPLETE
SAVE & EXIT
EXIT W/O SAVING
CALIBRATION
CLOCKWISE COMPLETE
REVERSE DIRECTION
FOR CCW
Attach necessary
weight fixtures.
THEN PRESS ENTER
CALIBRATION
CLOCKWISE
APPLY LOAD
1 OF 5
ENTER LOAD:
10.000 LBFIN
THEN PRESS ENTER