Cooper Instruments & Systems TT03C Digital Cap Torque Gage User Manual
Page 13

Series TT03 Digital Torque Gauges
User’s Guide
12
8. The next, and final, screen appears as follows:
Click “Finish”. The Mark-10 USB device is now installed and ready to use. The COM port number
assigned by Windows may be identified in Device Manager, or in the communication application being
used, such as MESURgauge or HyperTerminal.
9.2 Communication Settings
To set up communication settings, select USB Settings from the menu. The screen appears as follows:
Configure the baud rate and data format as required for the application. Default values are as follows:
Baud Rate:
9,600
Data Format: Numeric + units
Other communication settings are permanently set to the following:
Data Bits:
8
Stop
Bits:
1
Parity:
None
10 CALIBRATION
10.1 Initial Physical Setup
The sensor should be mounted vertically to a test stand or fixture rugged enough to withstand a load
equal to the full capacity of the instrument. Vertical orientation is preferable to avoid side loading, which
can affect the readings. Suitable certified calibration equipment is required, and caution should be taken
while handling such equipment.
USB SETTINGS
+ Baud Rate
+ Data Format